When I worked for Carpenter Technology Corporation in their research
center, we encountered several instances where production-lab chart grain-size
ratings were between 4 and 5 for a number of specimens on an order (these
orders required tests on 20 specimens from different bars). When we re-tested
them in the R&D center, we got similar chart ratings. When we actually
measured the grain size, however, all ratings were between 5 and 6 on the ASTM
E 112 scale. As the criterion for pass/fail was a grain size of 5 or finer
(higher), this bias was important. Consequently, at a subsequent ASTM E-4
committee meeting, I conducted a “round-robin” test. I took two specimens of
different ferritic stainless steels and etched them electrolytically with 60%
HNO
3 at about 1 V dc for 1 to 2 minutes, which revealed the grain
boundaries about as perfectly as possible. This was in the days of film
photography. I took 4 x 5-inch negatives of both specimens at various magnifications:
50, 100, 200 and 500X. I then made 8 x 10-inch enlargements of each negative. I
took these to the next E-4 meeting.
A number of ASTM E-4 members counted
intercepts using the three-circle grid and then counted the grains within a
test circle and intersecting the test circle on seven micrographs. Three were
at different magnifications for a ferritic stainless steel and four were at
different magnifications for another ferritic stainless steel. All images were
taken from the same region. The people did not calculate the grain size; they
only collected the raw data. Prior to that, they used a comparison chart –
plate I of E 112 – to estimate the grain size of each micrograph.
A few people later digitized the images and
measured the grain size with image analysis systems. Random grid placement was
used for the intercept method, but the template
contained five test circles for the planimetric method. So, the placement on the micrograph was not
completely random but forced. I calculated the grain sizes for each person’s
tests. I didn’t want them to do so because they might go back and change their
chart ratings.
This work revealed a distinct bias between
the chart ratings and actual measurements, both planimetric and intercept.
There was no bias at all between the two measurement methods – the
Jeffries planimetric method versus the Heyn-Hilliard-Abrams intercept method –
as shown in the three accompanying plots.